6-16 | Model 933S UV Analyzer
c.
Standard Range
Align the two larger-diameter holes in the Window Cover with the
two larger-diameter holes in the Measuring Cell and connect it to
the Window Cover with the (3) M4 x 35 screws. Tighten the screws
evenly.
Low Range
Hold the Measuring Cell with the sample fitting near the Aperture
Plate upward and connect it to the Window Cover with the (3)
M4 x 35 screws. Tighten the screws evenly.
9. Reconnect the sample tubing to the Measuring Cell.
10. Power up the analyzer and computer.
11. Allow the analyzer to warm up then perform a leak check on the
sample system fittings that were disconnected. See “Sample System
Leak Check” in Chapter 3.
Analyzers without Heated Cells:
If the leak check passes, continue with Step 12.
Analyzers with Heated Cells:
If the leak check passes, replace the Insulating Jacket or close the Oven
door to allow the Cell to warm up to normal operating temperature
and stabilize and then continue with Step 12.
12. Perform an Auto-Setup.
13. After the Auto-Setup is complete, change the
Flow Control
to ‘
3
’
(
zero
) and perform a manual Zero. Click
OK
then
Apply
. Do not save
the changes to EEPROM.
14. After the Zero is complete, change the
Flow Control
to ‘
0
’ (
auto
). Click
OK
then
Apply
. Save the changes to EEPROM.
15. If the analyzer uses
Measured
Pressure Compensation (optional),
return the setting to
Measured
mode. To do this, view the
Cell/Misc
Parameters
dialog box and enter ‘
0
’ in the
Default Cell Pres
(pressure
compensation) field.
Setup (tab)
Optical
Bench
<<Auto-Setup>>
Содержание 933S
Страница 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 34: ...2 10 Model 933S UV Analyzer...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 933S UV Analyzer...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Страница 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Страница 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Страница 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Страница 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Страница 187: ...Maintenance and Troubleshooting 6 11 Figure 6 2 Optical Bench Assembly with Standard Range Measuring Cell...
Страница 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Страница 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Страница 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Страница 229: ...Maintenance and Troubleshooting 6 53 Figure 6 14 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Страница 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 268: ...A 4 Model 933S UV Analyzer Sample Flow Diagram Plumbing Schematic North American Style WX 933NA 2...
Страница 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Страница 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Страница 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Страница 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Страница 285: ...Appendix A Drawings A 21 Inlet and Vent Flame Arrestor Locations Environmental Purge Option 100 1783 1A...
Страница 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...