6-38 | Model 933S UV Analyzer
Manipulating the PMT Level and PMT Balance
In certain situations (for example: degradation of PMTs or lamps, or the
optics are contaminated), it may be necessary to change the
P300
and
P301
jumper settings on the
Optical Bench
board. Changing the jumper
positions provides coarse gain adjustment of the
Reference
PMT and there-
fore changes the
PMT Level
and
PMT Balance
.
There are six jumper positions (four in
P300
and two in
P301
) – at least
one jumper must be in one of the six positions at all times. The jumper
combinations control the coarse supply voltage ranges of the two PMTs,
Reference
-path and
Measure
-path. The Optical Bench Auto-Setup proce-
dure performs fine adjustment to the PMT supply voltages at one jumper
combination. The jumper combinations affect the Auto-Setup in terms of
increasing or decreasing the resulting PMT Balance and, to a lesser de-
gree, the PMT Level.
Figure 6-11 describes the effect on the PMT Balance and PMT Gain in rela-
tion to different jumper positions.
To achieve an acceptable
PMT Balance
and
PMT Level
:
1. If jumpers have been reconfigured, perform an
Auto-Setup
.
If the PMT signals are outside their normal range of operation after
the Auto-Setup is complete (typical range is 4500–7800 mVDC),
adjust the
Lamp Max
setting slightly and start another Auto-Setup.
2. After an acceptable
PMT Balance
and
PMT Level
has been achieved,
change the
Flow Control
to ‘
3
’ (
zero
) and initiate a Manual Zero then
click
OK
. Do not save the changes to EEPROM. Allow the Zero gas to
flow for 2–3 minutes.
From the
Status
tab observe the
Calibration/Solenoid Status
field to
view the progress of the Zero calibration.
3. After the Zero is complete, change the
Flow Control
back to ‘
0
’ (
auto
).
Click
OK
then
Apply
. Save the changes to EEPROM.
4. From the
Status
tab, click
Conc. Alarm Soft Enable
On/Off
to
Enable
the Concentration Alarms then click
OK
.
NOTE
Setup (tab)
Optical
Bench
<<Auto-Setup>>
Setup (tab)
Gas
Calibration
<<Lamp Max>>
Setup (tab)
Gas Calibration
<<Flow
Control>>
Содержание 933S
Страница 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 34: ...2 10 Model 933S UV Analyzer...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 933S UV Analyzer...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Страница 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Страница 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Страница 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Страница 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Страница 187: ...Maintenance and Troubleshooting 6 11 Figure 6 2 Optical Bench Assembly with Standard Range Measuring Cell...
Страница 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Страница 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Страница 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Страница 229: ...Maintenance and Troubleshooting 6 53 Figure 6 14 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Страница 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 268: ...A 4 Model 933S UV Analyzer Sample Flow Diagram Plumbing Schematic North American Style WX 933NA 2...
Страница 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Страница 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Страница 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Страница 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Страница 285: ...Appendix A Drawings A 21 Inlet and Vent Flame Arrestor Locations Environmental Purge Option 100 1783 1A...
Страница 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...