6-4 | Model 933S UV Analyzer
Analyzer Preventive Maintenance Schedule
To reduce the occurrence of problems with the analyzer follow this sched-
ule, which lists the frequency of maintenance required when caring for
the analyzer. Since most analyzer problems originate within the sample
system, the primary objective of this schedule is proper care of the sample
system.
Frequency Component/Task
Every 6 Months
Zero Gas Cylinder
Check the Zero Gas cylinder and replace it if empty/nearly empty.
Every 6 Months
Sheltered Systems
If the analyzer is installed in a custom shelter, check the air filters
for the shelter purge and air conditioning systems, and replace if
necessary.
Depending on the environmental conditions of the site, more
frequent filter replacement may be necessary.
9–12 Months
Source Lamps
Replace the source lamps.
The lamps may require replacement sooner if analyzer alarms
are set (“
w: ALC out of range
”, “
w: PMT signal out of range
”,
“
w: Excessive zero drift
”, or “
f: Lamp Failure
”). See “Source
Lamp Replacement” and “Troubleshooting and Diagnostics” in this
chapter.
Every Year
Measuring Cell
Replace the o-rings (and Cell Windows, if required) and clean the
interior surfaces to remove any contaminants.
This schedule is a minimum requirement. If other conditions are
found to be present, as discussed in “Measuring Cell Maintenance,”
more frequent cleaning may be required.
Every Year *
Filterblock Assembly (Optional)
Replace the o-rings, Filter Cartridges, Membrane Filters, and
Flow Restrictors. Clean the interior surfaces to remove any
contaminants.
* In natural gas streams that are extremely dirty, the Flow
Restrictor in the first filter stage (and possibly the second filter
stage) may need to be changed more often because it removes
most of the liquids and particulate from the gas.
See “Natural Gas Filterblock Maintenance” in this chapter.
Every 2 Years
Column Block Assembly
Replace the Check Valves, Filter, Flow Restrictors, and o-rings.
Clean the interior surfaces of each section to remove contaminants.
The Column Assemblies need to be replaced only if they are
damaged due to liquids or severe contamination.
See “Column Block Maintenance” in this chapter.
Содержание 933S
Страница 24: ...1 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 34: ...2 10 Model 933S UV Analyzer...
Страница 35: ...Specifications 2 11...
Страница 36: ...2 12 Model 933S UV Analyzer 2016 12 16...
Страница 37: ...Specifications 2 13...
Страница 38: ...2 14 Model 933S UV Analyzer...
Страница 39: ...Specifications 2 15...
Страница 46: ...2 22 Model 933S UV Analyzer This page intentionally left blank...
Страница 52: ...3 6 Model 933S UV Analyzer Figure 3 1 Analyzer backpan dimensions European...
Страница 91: ...Installation and Start Up 3 45 Figure 3 24 Column Profile...
Страница 94: ...3 48 Model 933S UV Analyzer This page intentionally left blank...
Страница 176: ...5 12 Model 933S UV Analyzer This page intentionally left blank...
Страница 187: ...Maintenance and Troubleshooting 6 11 Figure 6 2 Optical Bench Assembly with Standard Range Measuring Cell...
Страница 190: ...6 14 Model 933S UV Analyzer Figure 6 3 1 Standard Range Cell 40 cm or smaller...
Страница 191: ...Maintenance and Troubleshooting 6 15 Figure 6 3 2 Low Range White Cell 81 2 cm P N 100 1780 1A...
Страница 197: ...Maintenance and Troubleshooting 6 21 Figure 6 4 Column Block assembly...
Страница 229: ...Maintenance and Troubleshooting 6 53 Figure 6 14 Optical Bench Chopper Assembly Maintenance diagram 100 2313 1A...
Страница 246: ...6 70 Model 933S UV Analyzer Figure 6 18 Microcontroller board P N 100 1781 SW400...
Страница 264: ...7 8 Model 933S UV Analyzer This page intentionally left blank...
Страница 268: ...A 4 Model 933S UV Analyzer Sample Flow Diagram Plumbing Schematic North American Style WX 933NA 2...
Страница 272: ...A 8 Model 933S UV Analyzer Lower Enclosure Layout WX 14838...
Страница 274: ...A 10 Model 933S UV Analyzer RS 232 Communications Cable Connections 300 9480...
Страница 279: ...Appendix A Drawings A 15 Optical Bench Board P N 100 1662...
Страница 280: ...A 16 Model 933S UV Analyzer PMT Buffer Board 100 0140...
Страница 285: ...Appendix A Drawings A 21 Inlet and Vent Flame Arrestor Locations Environmental Purge Option 100 1783 1A...
Страница 288: ...S 2 Model 933S UV Analyzer This page intentionally left blank...