Particulate Measurement System
User Guide
4750-PM
2
Technical Data
2.1 Process and Application Conditions
Stack temperature range
(at monitoring point)
-20 to 250 °C (-4 to 480 °F)
Option: up to 400°C (750 °F)
Stack pressure
±20 mb(g)
Stack diameter
B
Ø 1.5 to 15 m (5 to 50 ft.)
Sensor – Key Data
Measurement type
Dust concentration (in mg/m³)
Measurement capability
0 – 1000 mg/m³
Resolution
0.1 mg/m³
Zero drift
<3%
Span drift
<3%
Response time
2 seconds
Environmental
information
Application conditions
Suitable for measurement in non-condensing flue
gases.
Note: not recommended for use in applications
with water droplets
Location
This equipment is for outdoor or sheltered use.
2.2 4750-PM Sensor Specification
Sensor version
Standard sensor: 0 – 250 °C (up to 480 °F)
Weight
9.6 kg (21 lb)
Overall dimensions
L 354 x H 216 x B 204 mm (14 x 8.5 x 8 in.)
incl. flange and cable glands
Enclosure
Temperature range
-20 to 50 °C (-4 to 120 °F), ambient
C
Dimensions
W 199 x H 180 x D 107 mm (7.8 x 7 x 4.2 in.),
incl. glands and hinges
Protection rating
IP65
Material
Die-cast aluminium LM6M (epoxy coated)
Cable entries
3 x M20 cable glands
B
Include standoff length (see Section 4.4.2 for details).
C
Care should be taken to install the instrument out of direct sunlight to
minimize the effect of radiating heat, which may subject the sensor to
temperatures above +50 °C (120 °F).
2 - 0
Содержание 4750-PM
Страница 4: ...Particulate Measurement System User Guide 4750 PM Blank...
Страница 6: ...INTRODUCTION 1...
Страница 11: ...TECHNICAL DATA 2...
Страница 15: ...2 3 User Guide Particulate Measurement System 4750 PM...
Страница 18: ...PRODUCT DESCRIPTION 3...
Страница 24: ...INSTALLATION 4...
Страница 36: ...OPERATION 5...
Страница 43: ...5 7 User Guide Particulate Measurement System 4750 PM Fig 5 2 Engineering Menu...
Страница 53: ...MAINTENANCE 6...
Страница 59: ...RESETTING THE SENSOR 7...
Страница 61: ...GENERAL NOTICES 8...
Страница 63: ...4750 PM Particulate Measurement System User Guide...