5 - 13
User Guide
Particulate Measurement System
4750-PM
5.6.2 Self-check LED
The Self-check LED is used to indicate a sensor check fail (see item 3, Fig.
3-2). The
4750-PM
has manually activated sensor performance checks.
During normal operation, if the Self-Check LED turns RED, the Purge Air Flow
check has detected that the purge air flow has stopped or is insufficient.
Connection
Description
GREEN
(solid)
Sensor checks PASS
GREEN
(flashing)
Sensor checks RUNNING
RED
(solid)
A sensor check has failed (Relay 1 activates).
5.6.3 Viewing Sensor Check Results
For information on initiating sensor checks manually from the sensor refer
to Section 5.2.9. Refer to the Standard or Engineering menu maps (Sections
5.2 and 5.3 respectively).
1) If required, select the
DUST
menu, then press RIGHT repeatedly to
display each self-check and the current reading (see table below for
details).
2) Press RIGHT to return to the main menu or press LEFT to return to the
previous setting.
Sensor Check
Reading Displayed
ZERO
Zero test result
SPAN
Span (or Upscale) test result
FLOW
Purge air flow test result (if available)
5.7 Auditing the Sensor
The Manual Audit Unit (MAU) is a standalone unit that is used to audit the
instrument to ensure operational performance. The MAU is a single-value,
compact audit kit or block. The audit unit filter is calibrated to yield a
different instrument response; the value is labelled on the front of the MAU.
The MAU is attached to the sensor body opening (see Fig. 5-4) to check
the optical path and measure the instrument’s response to a stable scatter
source.
5.7.1 High-Range Audit Value Units
Audit units with a high-range audit value (ACCAUD-XXX-1-300) are marked
with a label (see Fig. 5-3). If an audit unit has been calibrated for a high-
range setting, ensure that the
LPUR
(laser power) setting is set to ‘ON’
before commencing the audit check.
NOTE
Laser power may already be set to ‘ON’ for normal operation.
Therefore, when the audit check is complete, always return LPUR
to its initial setting (see Section 5.4.4 for details).
Содержание 4750-PM
Страница 4: ...Particulate Measurement System User Guide 4750 PM Blank...
Страница 6: ...INTRODUCTION 1...
Страница 11: ...TECHNICAL DATA 2...
Страница 15: ...2 3 User Guide Particulate Measurement System 4750 PM...
Страница 18: ...PRODUCT DESCRIPTION 3...
Страница 24: ...INSTALLATION 4...
Страница 36: ...OPERATION 5...
Страница 43: ...5 7 User Guide Particulate Measurement System 4750 PM Fig 5 2 Engineering Menu...
Страница 53: ...MAINTENANCE 6...
Страница 59: ...RESETTING THE SENSOR 7...
Страница 61: ...GENERAL NOTICES 8...
Страница 63: ...4750 PM Particulate Measurement System User Guide...