3.5 Principle of Operation
Particles in the stack are illuminated by a laser and the amount of light
scattered back from the particles is measured. Stray scattering from the
far side of the stack wall is eliminated by tuning the instrument’s field-of-
view. Ambient light is eliminated by use of a modulated laser source. The
instrument response is proportional to the dust concentration, allowing the
instrument to be calibrated to provide an mg/m³ reading.
Fig. 3-4 Principle of Operation