The AMI Model 601 is designed to operate in an AMI power supply system consisting 

of a unipolar power supply, a power supply programmer, and a superconducting 

magnet. The following paragraphs document the procedures and interconnects 

necessary to use the Model 601.

WARNING

: If the Model 601 is used in a manner not specified by AMI, 

the protection provided by the equipment may be impaired.

A. Carefully remove the Model 601 from the shipping container and remove all packaging 

material.

NOTE

: If there is any shipping damage, save all packaging material 

and contact the shipping representative to a file a damage claim. Do not 

return the instrument to AMI unless prior authorization has been 

received.

B. If the Model 601 is to be mounted in a rack:

Install the Model 601 in a 19" wide instrument rack by securing the front panel to 

the rail in each of the four corners with mounting hardware supplied by the cabinet 

manufacturer.

WARNING

: The Model 601 must have sufficient cooling air to prevent 

damage from overheating. The cooling air inlets on the front panel and 

the cooling air outlets on the rear panel must not be obstructed.

C. Install the Model 601 in a magnet power supply system by completing the following 

steps:

CAUTION:

 Do not overtighten the nuts on the current lugs of the Model 

601 (maximum torque is 50 in-lbs). Overtightening can result in 

damage to the terminals.

1. Connect a power lead cable from the power supply positive (+) output lug to the 

positive (+) lug on the rear of the Model 601. No other connections should be made 

to the Model 601 positive (+) lug.

2. Connect a power lead cable to the negative (

) lug on the rear of the Model 601 to 

the positive (+) current lead of the superconducting magnet. No other connections 

should be made to the Model 601 negative (

) lug.

WARNING:

 Exercise care near the power lugs when operating a 

magnet. Metallic objects shorted across the lugs may conduct large DC 

Содержание 601

Страница 1: ...loss of AC power to a power supply system the Model 601 will ramp the system to zero current at a discharge rate of approximately 5 VDC II SPECIFICATIONS 25 C AC Input Ratings 100 to 240 VAC 10 50 60 Hz 100 VA max DC Input Ratings 5 VDC 5 4 A max Maximum Magnet Operating Current 130 A 25 C derated linearly to 100 A 40 C Nominal Discharge Voltage 5 0 VDC 2 Discharge Voltage Temperature Coefficient ...

Страница 2: ...front panel to the rail in each of the four corners with mounting hardware supplied by the cabinet manufacturer WARNING The Model 601 must have sufficient cooling air to prevent damage from overheating The cooling air inlets on the front panel and the cooling air outlets on the rear panel must not be obstructed C Install the Model 601 in a magnet power supply system by completing the following ste...

Страница 3: ...sitive and negative lugs will be ganged together as illustrated below using a conductive plate Simply follow the installation directions in Step C above by treating each ganged lug as a single output lug The power leads from the supply and to the magnet may be connected to any mounting lug if an additional lug is not provided in the ganged conductive plate In addition to the ganged lugs the coaxia...

Страница 4: ...is not energized C If the FAULT LED is energized then one or more of the internal energy absorbing elements has malfunctioned or power has been lost to the rear panel power connector An audible alarm will also sound when the FAULT LED is energized If the unit is in operation proceed immediately to a safe magnet system state typically zero current or a cooled persistent switch in a connected magnet...

Страница 5: ... the Model 601 and the AC power cord is also connected to the appropriate power receptacle 2 If the DC power converter is connected properly observe if an internal green LED is energized by looking through the rear panel grating If the green LED is energized then the DC power converter is operating correctly If the green LED is not energized then the DC power converter has failed or AC power to th...

Страница 6: ...er warranty and such repair or replacement shall constitute a fulfillment of all AMI liabilities with respect to its products All warranty repairs are F O B Oak Ridge Tennessee USA VIII RETURN AUTHORIZATION Items to be returned to AMI for repair warranty or otherwise require a return authorization number to ensure your order will receive the proper attention Please call an AMI representative at 86...