Rev. 6

4

AMI

 EXCELLENCE IN MAGNETICS AND CRYOGENICS

immediately

 upon connection to a power receptacle.

WARNING

: Do not position the external power supply or the Model 601 

so that it is difficult to disconnect the power cords.

The Model 601 DC power converter is designed to operate with 50/60 Hz power and 

voltages of 100 to 240 VAC ± 10%.

NOTE:

 If the Model 601 is operated without supplied power, the unit 

will function normally at currents above 0.5 A. Below 0.5 A, the Model 

601 will not develop the full discharge voltage of 5 volts if operated 

without power supplied to the rear-panel DC power converter connector. 

Operation without supplied power is only provided as a safety feature 

for situations where all facility power is lost to the magnet system. The 

Model 601 should not be routinely operated without supplied power.

IV.

OPERATION

A. During normal operation of the Model 601, no operator intervention is required.

B. The Model 601 is operating correctly if the FAULT LED is 

not

 energized.

C. If the FAULT LED is energized, then one or more of the internal energy absorbing 

elements has malfunctioned or power has been lost to the rear-panel power connector. 

An audible alarm will also sound when the FAULT LED is energized. If the unit is in 

operation, proceed immediately to a safe magnet system state (typically zero current or 

a cooled persistent switch in a connected magnet). Do not continue to operate the unit 

and refer to the 

Troubleshooting

 section below for further direction.

D. Remember to adjust the voltage limit settings of any power supply programmer to 

account for the additional voltage (5 V) required to operate the system with a Model 

601 installed.

WARNING

: The Model 601 must have sufficient cooling air to prevent 

damage from overheating. The cooling air inlets on the front panel and 

the cooling air outlets on the rear panel must not be obstructed.

V.

TROUBLESHOOTING

A.

The magnet system does not appear to charge

.

1. Check the system wiring and verify that the current flow direction through the 

Model 601 is from the positive (+) to the negative (

) lug.

2. Verify that the supplied external DC power converter is connected to the rear of the 

Содержание 601

Страница 1: ...loss of AC power to a power supply system the Model 601 will ramp the system to zero current at a discharge rate of approximately 5 VDC II SPECIFICATIONS 25 C AC Input Ratings 100 to 240 VAC 10 50 60 Hz 100 VA max DC Input Ratings 5 VDC 5 4 A max Maximum Magnet Operating Current 130 A 25 C derated linearly to 100 A 40 C Nominal Discharge Voltage 5 0 VDC 2 Discharge Voltage Temperature Coefficient ...

Страница 2: ...front panel to the rail in each of the four corners with mounting hardware supplied by the cabinet manufacturer WARNING The Model 601 must have sufficient cooling air to prevent damage from overheating The cooling air inlets on the front panel and the cooling air outlets on the rear panel must not be obstructed C Install the Model 601 in a magnet power supply system by completing the following ste...

Страница 3: ...sitive and negative lugs will be ganged together as illustrated below using a conductive plate Simply follow the installation directions in Step C above by treating each ganged lug as a single output lug The power leads from the supply and to the magnet may be connected to any mounting lug if an additional lug is not provided in the ganged conductive plate In addition to the ganged lugs the coaxia...

Страница 4: ...is not energized C If the FAULT LED is energized then one or more of the internal energy absorbing elements has malfunctioned or power has been lost to the rear panel power connector An audible alarm will also sound when the FAULT LED is energized If the unit is in operation proceed immediately to a safe magnet system state typically zero current or a cooled persistent switch in a connected magnet...

Страница 5: ... the Model 601 and the AC power cord is also connected to the appropriate power receptacle 2 If the DC power converter is connected properly observe if an internal green LED is energized by looking through the rear panel grating If the green LED is energized then the DC power converter is operating correctly If the green LED is not energized then the DC power converter has failed or AC power to th...

Страница 6: ...er warranty and such repair or replacement shall constitute a fulfillment of all AMI liabilities with respect to its products All warranty repairs are F O B Oak Ridge Tennessee USA VIII RETURN AUTHORIZATION Items to be returned to AMI for repair warranty or otherwise require a return authorization number to ensure your order will receive the proper attention Please call an AMI representative at 86...