Model 601, and the AC power cord is also connected to the appropriate power 

receptacle.

3. Check the voltage limit settings of any power supply programmer and verify the 

limits are set to a value greater than 5 volts plus the charging voltage and any 

power lead voltage drops.

B.

The system current ramps slowly from zero

.

If the Model 601 is used in conjunction with an AMI Model 430, then an initial 

charging delay will be observed when operating without an inductive load (e.g. a 

persistent switch is not heated on a connected magnet). To decrease the amount of 

time delay, increase the ramp rate to 1 A/sec or greater value. Remember to 

decrease the ramp rate, if necessary, before heating the persistent switch of the 

connected magnet and attempting to ramp the current. The charging delay will not 

be observed when operating with an inductive load.

C.

The FAULT LED is energized with audible alarm

.

1. Verify that the supplied external DC power converter is connected to the rear of the 

Model 601, and the AC power cord is also connected to the appropriate power 

receptacle.

2. If the DC power converter is connected properly, observe if an internal green LED 

is energized by looking through the rear-panel grating. If the green LED is 

energized, then the DC power converter is operating correctly.

If the green LED is not energized, then the DC power converter has failed, or 

AC power to the DC power converter has been lost. Contact an Authorized AMI 

Technical Support Representative for a replacement.

3. If the DC power converter is connected and the internal green LED is energized, 

then the FAULT LED indicates the failure of an internal energy absorbing 

element. Do not continue to operate the unit and contact an Authorized AMI 

Technical Support Representative for further instructions.

VI.

MAINTENANCE

The only routine maintenance required is to keep the exterior surfaces of the 

instrument clean by gently wiping with a damp cloth moistened with a mild detergent. 

The front and rear panel vents of the Model 601 should also be kept free of obstructions 

or excessive dust to allow for proper cooling of the unit.

VII. WARRANTY

All products manufactured by AMI are warranted to be free of defects in materials and 

Содержание 601

Страница 1: ...loss of AC power to a power supply system the Model 601 will ramp the system to zero current at a discharge rate of approximately 5 VDC II SPECIFICATIONS 25 C AC Input Ratings 100 to 240 VAC 10 50 60 Hz 100 VA max DC Input Ratings 5 VDC 5 4 A max Maximum Magnet Operating Current 130 A 25 C derated linearly to 100 A 40 C Nominal Discharge Voltage 5 0 VDC 2 Discharge Voltage Temperature Coefficient ...

Страница 2: ...front panel to the rail in each of the four corners with mounting hardware supplied by the cabinet manufacturer WARNING The Model 601 must have sufficient cooling air to prevent damage from overheating The cooling air inlets on the front panel and the cooling air outlets on the rear panel must not be obstructed C Install the Model 601 in a magnet power supply system by completing the following ste...

Страница 3: ...sitive and negative lugs will be ganged together as illustrated below using a conductive plate Simply follow the installation directions in Step C above by treating each ganged lug as a single output lug The power leads from the supply and to the magnet may be connected to any mounting lug if an additional lug is not provided in the ganged conductive plate In addition to the ganged lugs the coaxia...

Страница 4: ...is not energized C If the FAULT LED is energized then one or more of the internal energy absorbing elements has malfunctioned or power has been lost to the rear panel power connector An audible alarm will also sound when the FAULT LED is energized If the unit is in operation proceed immediately to a safe magnet system state typically zero current or a cooled persistent switch in a connected magnet...

Страница 5: ... the Model 601 and the AC power cord is also connected to the appropriate power receptacle 2 If the DC power converter is connected properly observe if an internal green LED is energized by looking through the rear panel grating If the green LED is energized then the DC power converter is operating correctly If the green LED is not energized then the DC power converter has failed or AC power to th...

Страница 6: ...er warranty and such repair or replacement shall constitute a fulfillment of all AMI liabilities with respect to its products All warranty repairs are F O B Oak Ridge Tennessee USA VIII RETURN AUTHORIZATION Items to be returned to AMI for repair warranty or otherwise require a return authorization number to ensure your order will receive the proper attention Please call an AMI representative at 86...

Отзывы: