AMI
AMI MODEL 601 ENERGY ABSORBER
INSTALLATION, OPERATION, AND
MAINTENANCE INSTRUCTIONS
I.
INTRODUCTION
The AMI Model 601 Energy Absorber is designed to provide a compact, fast-rampdown
option of up to 5 VDC for AMI power supply systems. The Model 601 is designed to
operate safely in the event of loss of facility AC power by drawing the necessary power
for internal cooling from the superconducting magnet. Upon loss of AC power to a
power supply system, the Model 601 will ramp the system to zero current at a
discharge rate of approximately 5 VDC.
II.
SPECIFICATIONS @ 25
°
C
AC Input Ratings: . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 to 240 VAC ± 10%, 50/60 Hz,
100 VA max
DC Input Ratings:. . . . . . . . . . . . . . . . . . . . . . . . . . . 5 VDC ± 5%, 4 A max
Maximum Magnet Operating Current: . . . . . . . . . . 130 A @ 25
°
C, derated linearly to
100 A @ 40
°
C
Nominal Discharge Voltage:. . . . . . . . . . . . . . . . . . . 5.0 VDC ± 2%
Discharge Voltage Temperature Coefficient: . . . . . ± 150 ppm/
°
C max
Internal Series Resistance:
1
. . . . . . . . . . . . . . . . . . . 2 m
± 1 m
Rated Operating Temperature: . . . . . . . . . . . . . . . . 0 to 40
°
C @ 100 A,
0 to 25
°
C @ 130 A
Rated Operating Relative Humidity:. . . . . . . . . . . . 0 to 95% non-condensing
Torque Limit on Current Terminals: . . . . . . . . . . . . 50 in-lbs.
1
(nominally adds 200 mV voltage drop at 100 A)
Instruction manual symbol: the product is marked with this symbol when it
is necessary for you to refer to the instruction manual in order to protect
against damage to the product or personal injury.