Symptom
Possible
Cause
How to Verify
Corrective Action
Buzzer or
Relays not
activating.
Bad Connection
to Buzzer or
Relays.
Configuration
incorrect.
No continuity between COM
and NO Bus Module
terminals.
No continuity between Relay
coil connection and relay
coils.
Re-seat wiring connections for
buzzer and relay coils.
Check UTx-> Buzzer menu
parameters. Check if either
Alarm or All Levels are
selected in item.
Check UTx-> Alarms-> Alarm
1/2/3/4-> Trigger menu
parameters; ensure Relay 1
and 2 are checked (enabled).
The Relay 1 coil is energized
during the WARN state, while
the Relay 2 coil is energized
during the ALARM state.
Alarm 1 is used to control
Relay 1 for CO.
Alarm 2 is used to control
Relay 2 for CO.
For 1BVC only:
Alarm 3 is used to control
Relay 1 for NO
2
.
Alarm 4 is used to control
Relay 2 for NO
2
.