5.3.4 CALIBRATION PROCEDURE
5.3.4.1 Calibration Menu
Navigate to the Service Menu and select the sensor to be calibrated. If the
ENTER
button is
pressed then the following menu is shown. This menu is common to all sensors therefore selecting
any sensor will produce the same menu.
The
UP
and
DOWN
buttons can be used to navigate the list of service menus for the selected
sensor. The sensor’s service menu provides the follo ing menu items
Table 5-17: Calibration Menu
Menu Item
Description
ADAPTiCal
Select this menu item to start calibration for the selected sensor
As Found
Allows user to enter the background concentration reading before performing
a bump test. (Refer to Section 5.3.3 for further explanation).
Window
Allows user to edit the custom window filter for the selected sensor (Refer to
Section 5.2.9)
Override
Allows user to edit the override the current sensor value for the selected
sensor. (Refer to Section 6.4.1)
Reset
Allows user to trigger a software reset of the selected sensor
Exit
Select this menu item to return to the main service menu. The current sensor
will be selected when the main service menu appears
ith ‘ADAPTi al’ highlighted as a ove , press
ENTER
to initiate a calibration sequence for the
selected sensor. Once initiated, the user will be presented with the following screens in which
the background and span level concentrations used for the calibration can be entered:
onte t sensitive help for the currentl highlighted menu item
A A Ti al
indo
tart A A Ti al
As ound
onte t sensitive help for the currentl highlighted menu item
A A Ti al
indo
tart A A Ti al
As ound