May 2020, Rev 4 

Page 

4

 of 

6

 

 

TROUBLE SHOOTING AND MAINTENANCE 

 

 

PART A.  

THE FLOW SYSTEM

 

 

A1. 

DIGITAL FLOW INDICATOR 
(i) 

NIL DISPLAY 

Check    :- 

Batteries are installed correctly. 

 

That lead wires from the Flowsensor are appropriately fastened to the 
 correct terminals. 

 

(ii) 

INDICATOR ERROR 

Check 

:-   Flowsensor operation, (See 

‘Flowsensor’ below). 

 

 
 
NOTES   

1. 

The digital flow indicator is individually factory calibrated & issued with a “K” factor to match the  
unit.  The “K” factor is programmed into the flow indicator during manufacture and is shown on a label on 
the side of the indicator.    

 

2. 

Page 5 & 6 of this manual show details on correct wiring & battery installation.  

3. 

Page 2 & 3 of this manual provides information on the normal operating display & also how to reset the 
resetable total (if desired). 

 

 
A2. 

THE FLOWSENSOR 

The operation of the flowsensor is dependant on correct installation in the installation fitting.  The flowsensor has 
two lugs inside the cap which locate in the slots in the flowsensor insert.  The insert slots must be parallel with the 
liquid flow.  Care must be taken that the flowsensor is not over-tightened in the insert, as this may cause the insert 
to rotate, leaving the paddlewheel misaligned to the flow direction. 
 
Debris in the flow may foul or damage the paddlewheel, which would then require removal and replacement. 

 

To remove the paddlewheel from its housing rotate th

e ‘red’ locking cap anti-clockwise until the cap is clear of the 

housing insert, then using the wire bale, remove the complete paddlewheel assembly from its housing b

y ‘gently’ 

twisting and pulling outwards. 

 

               

A3. 

PADDLEWHEEL REPLACEMENT 

 

(i) 

With a small flat-bladed screwdriver, gently pry away one of the paddlewheel mounting 

lugs from the pin.

   

(ii) 

When one end of the pin is free gently work the paddlewheel and pin out of the  

remaining lug.  

(iii) 

Clean or replace. 

(iv) 

To reinstall the paddlewheel and pin reverse steps (i) and (ii) above.

 

(v) 

The paddlewheel should spin freely when flicked by the finger. 

 

The flowsensor 

‘O-Rings’ should be checked periodically and replaced when necessary.   

The flowsensor shaft, 

‘O-Rings’ and insert barrel need lubrication as required. 

Lubricate with silicone oil.   

 
 
 
 
 
 
 
 
 
 
 
 