May 2020, Rev 4
Page
6
of
6
PART D.
BATTERY INSTALLATION AND REPLACEMENT
Two
3.6V Lithium thionyl Chloride
batteries, AA-size (7400-0011) are included with the Totalizer.
NOTE: THE 8150 WILL NOT OPERATE WITH STANDARD 1.5V ALKALINE BATTERIES. USE 3.6 V
LITHIUM BATTERIES ONLY
!
•
Observe Polarity!
Note that both batteries should face the same direction.
•
When the ‘low battery’ indicator appears on the display,
both
batteries should be
replaced within 90 days.
•
Remove and replace battery #1 first, then remove and replace battery #2.
This ensures that all settings and totalizer values are saved.
•
If the low battery symbol reappears for more than 10 seconds after installing new
batteries, one battery is reversed, or battery #2 was installed before battery #1.
•
Secure the batteries by fastening the hook-and-loop straps.
CAUTION
When replacing batteries
remove and replace
t h e battery at a time
Remove and replace
this battery first.
+ +
1
2
3
4
5
-
-
6
Note: Lithium Batteries
Dispose of properly!
Replace with 3.6 V Lithium battery
DISPOSE OF EXPENDED BATTERIES PROPERLY!
Lithium batteries contain hazardous chemicals.
Dispose of batteries according to local regulation.
AMBIT
Office & Factory | 42 Titan Drive, Carrum Downs, Victoria, 3201.
INSTRUMENTS PTY LTD
Ph: 03 9776 8888 | E: [email protected] | W: ambitinst.com.au
ABN 50 005 613 641