2000 Count Manual Ranging Digital Multimeter
Multímetro Digital de Rango Manual, 2000 Cuentas
Multimètre Numérique à Sélection de Gamme Manuelle, 

Écran (2 000 Comptes)
B07VY41YHM

Содержание B07VY41YHM

Страница 1: ...2000 Count Manual Ranging Digital Multimeter Mult metro Digital de Rango Manual 2000 Cuentas Multim tre Num rique S lection de Gamme Manuelle cran 2 000 Comptes B07VY41YHM...

Страница 2: ...English 3 Espa ol 28 Fran ais 56...

Страница 3: ...e following WARNING Risk of electric shock Improper use of this product can cause damage shock injury or death WARNING Risk of electric shock The safety features of this product may not protect the us...

Страница 4: ...probe tip shrouds after use For indoor use This meter is protected to CAT lll 600 V CAT ll 600 V Pollution Degree 2 This product must be used by trained users only Do not exceed the maximum allowable...

Страница 5: ...disconnect the connector leads from the circuit under test Inspect the condition of the connector leads and the product itself for any damage before operation Repair or replace any damage before use A...

Страница 6: ...ve the connector leads before replacing the battery or fuses Battery Warnings Always insert batteries correctly with regards to polarity and marked on the battery and the product Exhausted batteries s...

Страница 7: ...sult in minor or moderate injury NOTICE Indicates a practical tip advice or practice not related to personal injury This symbol adjacent to another symbol or terminal indicates the user must refer to...

Страница 8: ...om a Cat III source Outlets more than 60 ft 20 m from a Cat IV source CAT III 600V Product is designed to protect the user during measurements performed in the building installation up to 600 V Three...

Страница 9: ...This product may be used only under the conditions and for the purposes for which it was designed No liability will be accepted for damages resulting from improper use or non compliance with these in...

Страница 10: ...10 EN Product Description A B C D E G F...

Страница 11: ...ry compartment B HOLD button I 200 mA fuse C Function switch J 10 A fuse D 10A input jack K Tilt stand E COM input jack L Test probes with connector leads F backlight button M Probe tip shrouds G V mA...

Страница 12: ...EN 1 HOLD Display hold 2 Minus sign 3 Low battery indicator 4 Display reading Function switch 4 1 2 3 DC voltage Resistance AC voltage Diode test continuity Battery test DC current DC current 200 mA O...

Страница 13: ...en the low battery indicator 3 is shown on the display A Open the tilt stand K Loosen the screws of the battery compartment cover and remove it Insert a 9 V battery and attach it to the snap connector...

Страница 14: ...HOLD button B again to return to normal operation DC voltage measurement NOTICE Doing voltage measurements while The connector lead L is left on the 10A ampere input jack D The connector lead L is on...

Страница 15: ...voltage measurement mode Voltage Function switch C position 200 mV 200 mV 2000 mV 2000 mV 20 V 20 V 200 V 200 V 600 V 600 V Insert the black connector lead L into the COM input jack E Insert the red...

Страница 16: ...ctor lead L is on the V mA input jack G but the switch C is set to the DC current function It will blow the product s internal fuse if it draws a current greater than the internal fuses are rated for...

Страница 17: ...ent on circuits that exceed 250 volts NOTICE Do not make 10 A current measurements for longer than 30 seconds Exceeding 30 seconds may cause damage to the product and or the connector leads L For curr...

Страница 18: ...n line track and separate the 2 points where the flow of current must be measured Connect in series by attaching securely the test probes L each at the 2 open points of the track line Apply power to t...

Страница 19: ...k E Insert the red connector lead into the V mA input jack G Set the function switch C to the appropriate resistance measurement mode Resistance Function switch C position 200 200 2000 2000 20 k 20 k...

Страница 20: ...Forward voltage typically indicates 400 to 700 mV A shorted diode indicates near 0 V and an open diode indicates I in both polarities Continuity check WARNING Risk of electric shock To avoid electric...

Страница 21: ...st Set the function switch C to the 1 5 V BATT or 9 V BATT position Insert the black connector lead L into the COM input jack E Insert the red connector lead L into the V mA input jack G Place the con...

Страница 22: ...rs with a soft brush Never use corrosive detergents wire brushes abrasive scourers metal or sharp utensils to clean the product Replacing the fuses WARNING Risk of electric shock Disconnect the produc...

Страница 23: ...se s and install the new fuse s Close battery compartment H and tighten the screws Storage Store the product in its original packaging in a dry area Keep away from children and pets Specifications NOT...

Страница 24: ...z to 450 kHz sinusoidal 50 60 Hz all waves DC voltage Range Resolution Accuracy 200 mV 1 V 0 5 f m 2 digits 2000 mV 0 1 V 20 V 0 01 V 200 V 0 1 V 0 8 f m 2 digits 600 V 1 V DC current Range Resolution...

Страница 25: ...ts 2000 1 20 k 0 01 k 200 k 0 1 k 2000 k 1 k 1 f m 2 digits Test current 0 35 mA Continuity Measurement Output Test current max 1 mA Beeping tone when resistance is less than 30 Diode Test current Out...

Страница 26: ...V ceramic fast blow Battery test current 9 V 6 mA 1 5 V 100 mA Overrange indication I is displayed Display 2000 counts backlit LCD Measurement rate 2 readings per second nominal Input impedance 1 M V...

Страница 27: ...ousehold waste Take them to an appropriate disposal collection site Feedback and Help Love it Hate it Let us know with a customer review AmazonCommercial is committed to delivering customer driven pro...