9. Valve oven temperature:
The valve oven will get heated to this temperature value. It has to be in
the range of +40°C to +150°C.
10. Detector temperature:
The detector will get heated to this temperature value. It has to be in
the range of +130°C to +200°C for a FID detector or in the range of
+40°C to +150°C for a PID detector.
11. Hydrogen flow:
If you have a FID you can enter the hydrogen flow for the detector
here. The flow is regulated by a mass flow controller. The range is
from 25 ml/min to 75 ml/min. Standard is 40 ml/min.
12. Chrom Perfect method
In this menu item you can select the method for the chromatographic
analysis. See chapter 11.4.2.11 for more information about the Chrom
Perfect method.