Operating Manual GC 5000 BTX; Version 5 – 11/2010
139
14.4 Error messages
The GC 5000 BTX can show 8 different error messages which refer to
severe hardware errors within the analyzer.
Note!
Errors can be only maintained by the manufacturer. Therefore contact AMA
Instruments. Besides that first shut down the system and check the
communication cables. Disconnect the system from mains supply and
switch it on after 30 seconds.
Type of Error
Cause
Information
Error 1:
“Communication
breakdown to
temperature controller.
The program has been
stopped!”
Built-in computer gets no
connection to the temp-
erature controller.
This means that current temperature values are not
shown on the display anymore.
Heater set points cannot be changed.
Error 2:
“Communication
breakdown to valve
controller. The
program has been
stopped!”
Built-in computer gets no
connection to the valve
controller.
This means that current values for the carrier gas
pressure are not shown on the display anymore.
The valves are not switched anymore.
Error 7:
“Malfunction of the
detector temperature
sensor. The program
has been stopped!”
The temperature sensor is
broken.
All heaters are turned off.
Hydrogen supply is turned off.
The temperature sensor has to be exchanged.
Error 8:
“Malfunction of the
GC oven temperature
sensor. The program
has been stopped!”
The temperature sensor is
broken.
All heaters are turned off.
Hydrogen supply is turned off.
The temperature sensor has to be exchanged.
Error 9:
“Malfunction of the
valve oven
temperature sensor.
The program has
been stopped!”
The temperature sensor is
broken.
All heaters are turned off.
Hydrogen supply is turned off.
The temperature sensor has to be exchanged.
Error 10:
“Malfunction of the
enrichment stage
temperature sensor.
The program has
been stopped!”
The temperature sensor is
broken.
All heaters are turned off.
Hydrogen supply is turned off.
The enrichment tube has to be exchanged.
Содержание The GC 5000 BTX
Страница 1: ...Operating Manual Gas chromatograph GC 5000 BTX...
Страница 15: ...Operating Manual GC 5000 BTX Version 5 11 2010 13 Fig 2 Rear view Fig 3 Gas regulation inside the analyzer...
Страница 35: ...Operating Manual GC 5000 BTX Version 5 11 2010 33 Fig 5 Connection of analog outputs...
Страница 101: ...Operating Manual GC 5000 BTX Version 5 11 2010 99 Purging of sample line Sampling Waiting for GC Ready...
Страница 102: ...Operating Manual GC 5000 BTX Version 5 11 2010 100 Injection Waiting for end of cycle time...
Страница 104: ...Operating Manual GC 5000 BTX Version 5 11 2010 102 Purging of sample line Sampling Waiting for GC Ready...
Страница 105: ...Operating Manual GC 5000 BTX Version 5 11 2010 103 Injection Waiting for end of cycle time...