Operating Manual GC 5000 BTX; Version 5 – 11/2010
146
17.6 Control command ST
Telegram identification:
ST
Telegram length:
9
Telegram type:
Command
Brief description:
Output control words to a monitoring
instrument of the monitoring station.
Only one monitoring instrument is
addressed.
Field no.
Start byte
Data format
Field description
1
1
<STX>
Start of text
2
2
b
Length byte
3
3
ST
Telegram
identification
4
5
b
Monitoring instrument
ID
5
6
b
Control byte
6
7
<ETX>
End of text
7
8
<BCC1>
Upper nibble
8
9
<BCC2>
Lower nibble
Field number 4: Main instrument (component) controls the whole
monitoring station.
17.7 Control response TS
Telegram identification:
TS
Telegram length:
9
Telegram type:
Command
Brief description:
Acknowledgement of the control
command as response to ST.
Field no.
Start byte
Data format
Field description
1
1
<STX>
Start of text
2
2
b
Length byte
3
3
TS
Telegram
identification
4
5
b
Monitoring instrument
identification
5
6
b
Control byte
6
7
<ETX>
End of text
7
8
<BCC1>
Upper nibble
8
9
<BCC2>
Lower nibble
Содержание The GC 5000 BTX
Страница 1: ...Operating Manual Gas chromatograph GC 5000 BTX...
Страница 15: ...Operating Manual GC 5000 BTX Version 5 11 2010 13 Fig 2 Rear view Fig 3 Gas regulation inside the analyzer...
Страница 35: ...Operating Manual GC 5000 BTX Version 5 11 2010 33 Fig 5 Connection of analog outputs...
Страница 101: ...Operating Manual GC 5000 BTX Version 5 11 2010 99 Purging of sample line Sampling Waiting for GC Ready...
Страница 102: ...Operating Manual GC 5000 BTX Version 5 11 2010 100 Injection Waiting for end of cycle time...
Страница 104: ...Operating Manual GC 5000 BTX Version 5 11 2010 102 Purging of sample line Sampling Waiting for GC Ready...
Страница 105: ...Operating Manual GC 5000 BTX Version 5 11 2010 103 Injection Waiting for end of cycle time...