7.2 How to adjust the retention time windows
If you find 0.00 results although the peaks are present in the
chromatogram, the retention time windows are not at the correct position
and have to be adjusted.
1.
Menu bar
Parameters
CP Analysis
Load the last chromatogram in the CP Analysis module
Note
: You should have recorded at least three chromatograms to make
sure that the system runs stable now. This means as well that the first
run should be discarded.
You can load several chromatograms at the same time:
Select them
in
File open dialog
with
Ctrl/Shift
and mouse-click before clicking
“Open”
.
2.
Menu bar
Compare
Overlay
Mark „Overlay“ in the menu „Compare“.
Overlay of all chromatograms: They are listed on top of each other
instead of each one in a separate window.
How to load a
chromatogram
Содержание The GC 5000 BTX
Страница 1: ...Operating Manual Gas chromatograph GC 5000 BTX...
Страница 15: ...Operating Manual GC 5000 BTX Version 5 11 2010 13 Fig 2 Rear view Fig 3 Gas regulation inside the analyzer...
Страница 35: ...Operating Manual GC 5000 BTX Version 5 11 2010 33 Fig 5 Connection of analog outputs...
Страница 101: ...Operating Manual GC 5000 BTX Version 5 11 2010 99 Purging of sample line Sampling Waiting for GC Ready...
Страница 102: ...Operating Manual GC 5000 BTX Version 5 11 2010 100 Injection Waiting for end of cycle time...
Страница 104: ...Operating Manual GC 5000 BTX Version 5 11 2010 102 Purging of sample line Sampling Waiting for GC Ready...
Страница 105: ...Operating Manual GC 5000 BTX Version 5 11 2010 103 Injection Waiting for end of cycle time...