-28
-
H
PoorSig Detected
(1)Poor signal detected
(2)Transducers
installed
improperly
(3)Too much fouling
(4)The pipe liners are too
thick.
(5)Problem with transducers
cords
(1)Relocate measuring place
(2)Clean the spot
(3)Check the cords
(4)Check the coupler
Q
Frequ OutputOver
The actual frequency for the
Frequency Output is out of
the range set by the user
Check the value entered at
M66,M67,M68 and M69, and
try to enter a larger value on
M69
F
System RAM Error
Date Time Error
CPU or IRQ Error
ROM Parity Error
(1) Temporary problems
with RAM, RTC
(2) Permanent problems
with hardware
(1) power on again
(2) contact factory
1
2
3
Adjusting Gain
Instrument is in the progress
of adjusting the gain for the
signal, and the number
indicates the progressive
steps
K
Empty pipe
(1)
No liquid inside pipe
(2)
Setup error on M29
(1)
Relocate where the pipe
is full of liquid
(2)
Enter 0 on M29
§5.3 Other Problems and Solutions
(1)
When the actual flow inside the pipe is not standstill, but the instrument displays 0.0000 for
the flow rate, and 'R' displaying signal strength and the signal quality Q (value) has a
satisfactory value?
The problems are likely caused by the user who has used the `Set Zero' function on this
non-standstill flowing pipe. To solve this problem, use the `Reset Zero' function on menu
window M43.
(2) The displayed flow rate is much lower or much higher than the actual flow rate in the pipe
under normal working conditions.
(a)
There is probably an offset value wrongly entered by the user in M44. Enter '0' in
M44.
(b)
Problem with transducer installation.
(c) There is a 'Zero Point'. Try to 'zero' the instrument by using M42 and make sure that
the flow inside the pipe should be standstill.