-27-
5.Troubleshooting
§5.1 Power-on Error Displays and Counter-Measures
The Q-FP 201 ultrasonic flow meter provides an automatic power-on diagnosis for the hardware
problems. When any message (with the power on) in the following table displays, counter-measures
should be taken.
Error message
Causes
Counter-measures
ROM Testing Error
Segment Test Error
Problem with the software
(1)Power on again
(2)Contact with factory
Stored Data Error
The parameters entered by the
user lose integration.
When this message displays, the
user should press ENT key, and all
the configuration will be restored to
the default state.
Timer Slow Error
Timer Fast Error
Problem with the timer-keeper or
the crystal oscillator.
(1)Power on again
(2)Contact with factory
Date Time Error
Number errors with the calendar
Initialize the calendar by menu
window M61
Reboot repetitively
Hardware problems
Contact the factory
§5.2 Error Code and Counter-Measures
The Q-fp 201
ltrasonic flow meter will show Error Code in the lower right corner with a single
letter like I, R etc. on menu windows MOO, MO1, MO2, M03, M90 and M08. When any abnormal
Error Code shows, counter-measures should be taken.
Error
code
Correspondent
Message displayed
on MOS
Causes
Counter-measures
R
System Normal
No error
I
Detect No Signal
(1)No Signals detected
(2)Transducers
installed
improperly
(3)Too much fouling
(4)Pipe liners are too thick.
(5)Transducer cords are not
properly connected
(1)Relocate
measuring
location
(2)Clean the spot
(3)Check the cords
J
Hardware Error
Hardware problem
Contact the factory
www.automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111) 4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter