-32-
(I) Prefix W
The prefix W should be used in the network environment. The usage format is W + digit string
which stands for the IDN + basic command.
The digit string should have a value between 0 and 65534 except 13(0DI-0, 10 (OAH), 42(2AH,*),
38(261-1, &).
For example, if the IDN=12345 instrument is addressed and returning the velocity of that
instrument is requested, the command will be W12345DV(CR).
(2) Prefix N
The prefix N is a single byte IDN network prefix, not recommended in a new design. It is reserved
only for the purpose of the compatibility with the former versions
(3) Command Connector &
The
&
command connector can connect up to 6 basic commands to form a longer command so that
it will make the programming much easier.
For example, assume that the measurement of an instrument with IDN=4321 are going to be
returned, and (then) all the following 3 values--- (1) flow rate (2) velocity (3)POS totalizer---will be
returned simultaneously. The combined command would be W432IDQD&DV&DI+(CR), and the
result would be:
d-1.12m3/d(CR)
+3 .100m/s(CR)
+ 0m3(CR)
§6.5 Codes for the Keypad
The codes for the keypad should be used when the instrument is connected with other terminals that
operate the instrument by transmitting the 'NV command along with the keypad code. By this
function, remote operation of this instrument can be realized, even via the Internet.
Key
Hexadecimal
Key code
Decimal
Key code
ASCII
Code
0
30H
48
0
I
3 tH
49
1
2
32H
50
2
3
33H
51
3
4
34H
52
4
5
35H
53
5
6
36H
54
6
7
37H
55
7
Key
Hexadecimal
Key code
Decimal
Key code
ASCII
Code
8
381-1
56
8
9
391-1
57
9
3AH
58
•
3BH2OBH
59
MENU
3CH2OCH
60
<
ENT
3D1-1,0DH
61
♦
/+
3EH
62
>
♦
/-
3FH
63
?
www.automat.com.ar
Dr. R. Bielsa 175 (ex Morlote) C1427AZA Buenos Aires, Argentina. Tel.: (54111) 4555-0055 / Fax: (5411) 4555-0044
INSTRU Q-FP 201
Handheld Ultrasonic Flowmeter