-18-
M42
Zero point setup under the condition when there is no liquid running inside
the pipe.
M43
Clear the zero point set by the user, and restore the zero point set by the
factory
M44
Set up a manual flow bias. Generally this value should be 0.
M45
Scale factor for the instrument. The default value is '1'.
Keep this value as '1', when no user calibration has been made.
M46
Network environment Identification Number. Any integer can be entered
except 13(ODH, carriage return), 10 (OAH, line feeding), 42 (2AH), 38, 65535.
Every set of the instrument in a network environment should have a unique
IDN. Please refer to the chapter for communication.
M47
System locker to avoid modification of the parameters
M48
Not used
M49
Communication tester
M50
"Option" selection for the built-in logger. It also functions as the switch of
logger
M51
Time setup for the data logger
M52
(1) Data logging direction control. If `To RS-232' is selected, all the data
produced by the data logger will be transmitted out through the RS-232
interface
(2) If `To buffer ' is selected, the data will be stored into the built-in logger
memory
(3) Buffer transferring and buffer clearing
M53
Logger buffer viewer. It functions as a file editor. Use Dot, backspace UP and
DN keys to browse the buffer.
If the logger is ON, the viewer will automatically refresh once new data are
stored
M54
Not used
M55
Nod used
M56
Not used
M57
Not used
M58
Not used
M59
Not used
M60
99-year calendar. Press ENT for modification. Use the dot key to skip the
digits that need no adjusting.
M61
Display Version information and Electronic Serial Number (ESN) that are
unique for each Q-FP 201 series flow meter.
The users can employ the ESN for instrumentation management
M62
RS-232 setup. Baud rate can be 75 to 115200 bps