BYK-Gardner GmbH
Operation | 4
301 200 141 E 2010
wave-scan 3 ROBOTIC
23
4.3.3 Device Test
With USB connection and power supply the device can be tested. During
the test a measurement is performed. The wave-scan ROBOTIC performs a
non-contact measurement. A possible test setup is shown below.
1
2
3
4
5
Illustration 20:
Setup for device test
1 Sample to be measured
2 Measurement distance 15 mm
Measurement angel 2°
3 Instrument
wave-scan 3 ROBOTIC
4 Power cable
5 USB cable
Go to the tab
Device Test
.
Illustration 21:
Device test in smart-robotic
Perform following steps:
1.
Initialize
: Load device driver.
2.
Measure
: Get measurement data.
3.
De-Initialize
: Unload device driver.
With static sample only the dullness can be measured. To measure the val-
ues for the wavelength slowly move the sample (or the instrument) during
measurement.