4 | Operation
BYK-Gardner GmbH
22
wave-scan 3 ROBOTIC
301 200 141 E 2010
4.3.2 Device Connection
To access the device via
smart-robotic
, the instrument has be connected to
the power supply too.
1
2
3
Illustration 18:
Cabling for connection test
1 Instrument
wave-scan 3 ROBOTIC
2 USB port on instrument
3 USB port on PC
With power supply and USB connection the instrument can now be ac-
cessed in
smart-robotic
in the
Devices
>
Connection
tab.
Illustration 19:
Connection settings in smart-robotic
Perform following steps:
1. Activate the option
Work with USB
.
2. Select the
USB identifier
from the list.
3. Click the button
Check connection
.
4. Check if the status is
Connected
.
With an active USB connection the IP data – used in the productive environ-
ment for the LAN connection – can be sent to the device.