BYK-Gardner GmbH
Appendix | 6
301 200 141 E 2010
wave-scan 3 ROBOTIC
31
6.5 Technical Data
6.5.1 Measuring Data
Light source
Laser Diode LED and IR-SLD
Laser diode
Laser Class 1, P ≤ 390 µW,
λ
=670 nm,
DIN EN 60825-1:2015-07
Measure range
Dullness: 1 to 40
LW, SW: 0 to 100
Wa – We: 0 to 100
Structure spectrum
Dullness: < 0.1 mm
Wa: 0.1 to 0.3 mm
Wb: 0.3 to 1.0 mm
Wc: 1.0 to 3.0 mm
Wd: 3.0 to 10.0 mm
We: 10.0 to 30.0 mm
Scan length
5 / 10 / 20 cm
Scan speed
50 to 150 mm/sec
Repeatability
du < 40: 4% or > 0.4 (standard deviation)
du > 40: 6% or > 0.6 (standard deviation)
Reproducibility
du < 40: 6% or > 0.6 (standard deviation)
du > 40: 8% or > 0.8 (standard deviation)
Distance to surface
15 ± 2mm
Angle to perpendicu-
lar
± 2°
Object radius
> 500 mm
Smallest sample size
35 mm x 150 mm
Resolution
375 pixel per cm
Memory capacity
n/a – immediate data transfer
6.5.2 General Data
Temperature range
Storage: 0° C to 60° C
Operation: 10° C to 40° C
Relative humidity
Up to 85% non-condensing at 35° C
Operation altitude
Up to 2000 m (6561 ft)
Dimensions (LxWxH)
112 x 59 x 131 mm (4.4 x 2.4 x 4.8 in)
Weight
Ca. 550 g (1.21 lbs)
Interface
Initialization: USB-C (USB 2.0)
Operation: Ethernet LAN (M12-A)
Power supply (input)
Initialization: 5 VDC ; max. 1.75 A (USB-C)
Operation: 24 VDC ; max. 0.75 (M12-A)