Safety
•
Never allow untrained personnel to operate the system.
•
Never press the trigger or footswitch unless the handpiece is safely oriented at a
specific and intended target.
•
Never leave the system turned on, open or unattended during system
maintenance.
2.3.3. Burn
Hazards
The Accent XL system is a state-of-the-art radiofrequency system for medical,
aesthetic and cosmetic applications. The system emits radiofrequency energy that
can damage the skin in case of the use of excessive energy.
2.3.4.
Explosion and Fire Hazards
•
The system is not suitable for use in the presence of flammable mixtures with
air or oxygen.
•
Do not operate in the presence of volatile solvents such as alcohol, gasoline or
other solvents.
•
Do not use any flammable substances such as alcohol or acetone in the
preparation of the skin for treatment. If necessary, use soap and water to clean
before treatment.
•
If alcohol is used to clean and disinfect any part of the Accent XL system,
allow it to dry thoroughly before operating the system.
•
Flammable materials must be kept at a safe distance from the system.
2.3.5.
High Voltage Hazards
The system utilizes 110-120 VAC. To avoid personnel injury, do not operate the
system before ensuring that the exterior panels are properly closed. Do not attempt
to remove or disassemble the exterior panels.
The Accent XL system produces very high voltages in various components. Some
components may retain a charge after the power supply has been turned off, so no
part of the exterior housing should be removed, except by Alma Lasers authorized
personnel.
Whenever system maintenance is performed, never leave the Accent XL system
turned on, open or unattended.
2-6 Accent XL
™
Operator's Manual