Maintenance
7.4. Routine and Preventive Maintenance
The following routine maintenance procedures should be performed by the clinic
staff on a regular basis determined by the clinic protocol.
Warning
•
The Accent XL system generates high voltages and pulsed light/laser
radiation when powered up.
•
Always turn the system off and unplug the power cable before performing
maintenance procedures.
•
The interior of the system or its components may be serviced only by Alma
Lasers authorized technical personnel.
7.4.1.
Cleaning and Disinfecting the System
The outer surface of the system may be wiped clean with a soft cotton cloth
swabbed in 70% alcohol.
7.4.2.
Cleaning and Disinfecting the Handpieces
Caution
The coatings of the handpiece applicator tips are very delicate. Take all
necessary precautions in order to prevent scratching the tips.
The Accent XL handpieces, in particularly the applicator tips, are cooled
intensively when the system is in
Ready
mode. Water may condense on the
applicator tip's surface which could affect the treatment process and, under certain
conditions, may produce sparks. Avoid water condensation on the applicator tip's
surface. Wipe the surface of the applicator tip clean immediately before a treatment.
Following every treatment wipe the handpiece with a soft cotton swab dipped in
70% alcohol.
On a weekly basis check the handpiece tips for cleanliness. If necessary clean the
handpiece tips with a soft cotton swab dipped in 70% alcohol.
Caution
•
Do not wet the handpiece with water or immerse the handpiece in any
liquid.
•
During treatment use only the treatment oil recommended by Alma Lasers.
7-4 Accent XL
™