METRO POWERPOD™ SYSTEM —
INSTRUCTIONS FOR USE
6
www.metro.com
L01-622
REV 06/19
DANGER:
Metro PowerPod™ Systems are not for use in hazardous (classified) locations. Do not use or recharge the power
supply battery in oxygen enriched areas, areas where flammable anesthetics are used or stored, or any other
hazardous, classified location.
Shipment of Power Supply
Depending on the battery chemistry, the preparation and marking of the shipment are different.
Lithium Iron Phosphate Battery (Li-Fe)
The shipping container must have the following marking:
Class 9 Shipment:
Outside the USA in any shipping mode and air or sea transport inside the US, the power supply must be shipped as fully
regulated Class 9 (Hazardous Material / Dangerous Goods).
In this shipping mode the batteries can be connected or disconnected. The shipper must be certified for Li-Ion Class 9
shipments.
The power supply must be shipped per UN3481 Lithium ion batteries contained in equipment on cargo aircraft only.