Copyright © 2020
Alazar Technologies Inc.
ATS9353 User Manual
65
Additional Low-Frequency Analog Input
Bandwidth (-3 dB)
DC-coupled, 50
Ω
DC - 100 kHz
Maximum sample rate
200 KS/s single shot
Full scale input range
-0.5 V to +2.5 V, fixed
DC accuracy
±2% of full scale in all ranges
Input coupling
DC
Input impedance
50
Ω
±1
%
Input protection
±3 V
Absolute max. amplitude 5 V
p-p
Acquisition Memory System
Memory size
128 Million samples per channel
Record length
Software-selectable with 32-point resolution. Record length
must be a minimum of 256 points. There is no upper limit on
the maximum record length.
Number of records
Software-selectable from a minimum of 1 to a maximum of
infinite number of records
Pre-trigger depth
From 0 to 4080 for single channel
From 0 to 2040 for dual channel
Post-trigger depth
Record Length – Pre-Trigger Depth
Timebase System
Timebase options
Internal Clock or External Clock
Internal sample rates
500 MS/s, 250 MS/s, 100 MS/s, 50 MS/s, 20 MS/s, 10 MS/s,
5 MS/s, 2 MS/s, 1 MS/s, 500 KS/s, 200 KS/s, 100 KS/s,
50 KS/s, 20 KS/s, 10 KS/s, 5 KS/s, 2 KS/s, 1 KS/s
Internal clock accuracy
±2 ppm