MAINTENANCE
55
MAINTENANCE
Optomodule
1. Place the instrument in the upright position.
2.
Gently
clean the optomodule using a lint-free swab
dampened with the appropriate cleaning solution, as
specified in the “Cleaning” section. The cleaning solution
may be sprayed on difficult-to-remove residue to help wet
and soften the residue for easier removal.
3. After removing residue,
gently
rinse with a lint-free swab
dampened with water. Water may be sprayed on the cleaned
surfaces to rinse areas that are difficult to reach with a swab.
4.
Gently
dry with a lint-free swab or allow to air dry.
Valve Actuator
1. Place the instrument in the upright position.
2.
Gently
clean the valve actuator and actuator seal using a
lint-free swab dampened with the appropriate cleaning
solution, as specified in the “Cleaning” section. The cleaning
solution may be sprayed on difficult to remove residue to
help wet and soften the residue for easier removal.
3. After removing residue,
gently
rinse with a lint-free swab
dampened with water. Water may be sprayed on the cleaned
surfaces to rinse areas that are difficult to reach with a swab.
4.
Gently
dry with a lint-free swab or allow to air dry.
5. After cleaning, inspect the exposed tips of the valve
actuators. A broken tip may be supported by the actuator
seal and not appear defective. Lightly attempt to push the
tips of the valve actuators from side to side with a dry lint-
free swab. If a tip is not rigid, then it is broken and must be
replaced before using the instrument.
After cleaning
Inspect the exposed tips of the valve actuators for damage by
lightly pushing the tips of the valve actuators side-to-side with a
dry swab. If a tip is not rigid, it is broken and must be replaced
before using the pump.
To Clean (Continued)
Do not use isopropyl alcohol on
the optomodule.
CAUTION
Failure to perform these
inspections may result in
improper instrument operation.
WARNING
A
Optomodule
Valve
Actuator
Do not use isopropyl alcohol to
clean the valve actuators.
CAUTION
Care must be taken when
cleaning the vicinity of the valve
actuators to avoid damage and
breakage of the actuator tips.
Damage or breakage of the
actuator tips could cause an
uncontrolled flow condition.
WARNING
VALVE ACTUATOR
A
Содержание MedSystem III
Страница 37: ...34 GETTING STARTED T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Страница 45: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 42 ALARMS ADVISORIES AND PROMPTS...
Страница 61: ...58 MAINTENANCE T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K...
Страница 63: ...T H I S PA G E I N T E N T I O N A L LY L E F T B L A N K 60 MAINTENANCE...