Page 41
14
Technical Specifications
General specifications apply for the temperature range 5°C to 40°C. Accuracy specifications apply
for the temperature range 18°C to 28°C after 15 minutes warm-up with no load and calibration at
23°C. Typical specifications are determined by design and are not guaranteed.
OUTPUT SPECIFICATIONS (Each Output)
Voltage/Current
Ranges:
O/P1: 0V to 35V/0.1mA to 3A or 0V to16V/0.1mA to 6A
or 0V to 35V/0.1mA to 6A with O/Ps disabled.
O/P2: 0V to 35V/0.1mA to 3A or 0V to16V/0.1mA to 6A
or 0V to 35V/0.1mA to 6A with O/Ps disabled.
O/P3: 0V to 35V/0.1mA to 3A, 0V to 70V/0.1mA to 1.5A
or 0V to 70V/0.1mA to 3A with O/Ps disabled
O/P4: 0V to 35V/0.1mA to 3A, 0V to 70V/0.1mA to 1.5A
or 0V to 70V/0.1mA to 3A with O/Ps disabled
Output Power:
Up to 420W total output power
Voltage Setting:
Resolution: 1mV O/P1 ,O/P2; O/P3 & O/P4(35V) 10mV O/P3 &
OP4(70V)
Accuracy: O/P1,OP2,OP3 & O/P4(35V): ± (0.05% of s 3mV),
OP3 & O/P4(70V): ± (0.1% of s 10mV)
Current Setting:
Resolution: 0.1mA
Accuracy: ± (0·3% + 3mA) to 3A, ± (0·5% + 3mA) to 6A
Operating Mode:
Constant voltage or constant current with automatic cross-over.
CV or CC mode indication in display.
Output Switch:
Independent electronic switching with ON indication. In addition, Multi-On
and Multi-Off keys permit the outputs to be switched on/off synchronously.
Multi-On/Multi-Off
Action:
Individually settable delay between pressing of the Multi-On or Multi-Off key
and the turning on or off of the respective output. Delays settable between
10ms and 20 seconds. Separate delays for On and Off can be set.
Outputs can also be omitted from Multi-On or Multi-Off control.
Output Terminals:
Universal 4mm safety binding posts on 19mm (0·75”) spacing for Output;
screwless terminals for Sense.
MX100QP also has duplicate power and sense terminals at rear.
Ripple & Noise
(20MHz bandwidth):
Outputs 1 & 2 loaded at 16V/6A, Outputs 3 & 4 loaded at 35V/3A, CV mode:
All Outputs typically <0.5mVrms, <5mV pk-pk; 1mVrms max.
Rear terminals: 10mV pk-pk max. 1.5mVrms max.
O/P3 & O/P4 on 70V/3A range:
Typically <1mVrms, <10mV pk-pk; 1.5mVrms max.
Rear terminals: 15mV pk-pk max.
Load Regulation:
For any load change, measured at the output terminals, using remote sense:
Constant voltage: <0.01% ± 5mV
Constant current: < 0.01% ±0.5mA
Line Regulation:
Change in output for a 10% line change:
Constant voltage: <0.01% ± 5mV
Constant current: < 0.01% ± 250uA
Transient Response:
To within 50mV of set level for a 5% to 95% load change:
Front terminals: <150us.
Rear terminals: <175us
<500us (Outputs 1 & 2: 35V/6A range).
Temp. Coefficient:
Typically <100ppm/°C