2. PUTTING INTO SERVICE
1. Connect transducer to sockets M0 to M2 (2); (see Section 4).
2. Ensure that the power supply is provided via 9-volt battery or mains
adapter; (see Sections 3.1, 3.2).
3. To switch ON, push slide switch (1) located on left side of device upwards;
(see Section 3.3).
4. Displaying the measured values
Select MEASURED VALUE function by pressing the key
M
▲
(6).
Select measuring channels by pressing key
M
▲
and read out measured
values; (see Section 6.1.1).
5. Connect the cyclic measured value output to the printer or to the
computer. Connect the peripheral device via interface cable to socket A1;
(s. Manual, Sec. 5.2). On the peripheral device set 9600 bauds, 8 data bits,
1 stop bit, no parity; (see Section 9.2). If necessary enter the current
date and time-of-day; (see Section 9.1). Select
CYCLE
function ´
ZY
´ by
pressing the
FUNCTION
key and program the cycle; (see Section 6.3.2).
Select the output format in the
BAUD RATE
function ´
BR
´ by pressing keys
PROG,
▲
..,
PROG, ESC
; (see Section 6.3.3).
Start and stop cyclic measuring point scanning by pressing the
START/STOP
key; (see Section 6.3.2).
6. Saving measured values (only possible with option S)
Select
CYCLE
function ´
ZY
´ by pressing the
FUNCTION
key and program the
cycle; (see Section 6.3.2).
If necessary enter the current date and time-of-day; (see Section 9.1).
Start and stop measured value recording by pressing the
START/STOP
key;
(see Section 6.3.2).
Output data from memory to the printer or to the computer.
Connect the peripheral device via interface cable to socket A1; (see
Manual, Section 5.2). On the peripheral device set 9600 bauds, 8 data bits,
1 stop bit, no parity; (see Section 9.2).
Select the output format in the
BAUD RATE
function ´
BR
´ by pressing the keys
PROG,
▲
..,
PROG, ESC
; (see Section 6.3.3).
In the function ´
SF
´ output measured values by pressing the
MANU/PRINT
key; (see Section 6.5.3).
7. Evaluating a measuring operation
Select the function
MAXWERT
´
MH
´ or
MINWERT
´
ML
´ by pressing the
FUNCTION
key and call up the maximum and minimum values; (see Section 6.1.3).
ALMEMO
2390-5
11
Putting into service