6.4.7 Averaging over several measuring points
To determine the average value over several measuring points from Mb2 to
Mb1 at the same time, function channel ´M(n)´ is available. By programming
this function channel (e.g. as 2nd channel in the 3rd connector, measuring point
M12 in connector M2) one obtains the average value for all three measured va-
lues (from M00 to M02) because by default measuring point M00 is used as re-
ference channel Mb2 and the first measuring point in the connector of the
function channel is used as reference channel Mb1. For this purpose avera-
ging mode is not necessary.
6.5 Measured value memory (option S)
Measuring instrument ALMEMO 2390-5 is also available with the option “S” as
a data logger. The basic information on saving data in ALMEMO
®
devices is
provided in the Manual, Section 6.9. The internal data memory on the ALME-
MO 2390-5 is a 32-KB EEPROM with capacity sufficient for 3700 to 6000 mea-
sured values (depending on the channel number). In the event of a failure in
the supply voltage the measured data is retained; only the time-of-day is lost.
The EEPROM can be configured either as linear memory or ring memory; (see
Manual, Section 6.10.13.2).
Memory capacity display
If your measuring instrument has the data logger option (OA 2390-5S), you can,
by pressing the FUNCTION key, select the Memory free function ´MF´ and see
the remaining memory capacity in kilobytes. If the memory is full, the display
will read 0.0; if the memory is configured in linear form, no further measured va-
lues will be saved; if it is configured as a ring memory, old values will be over-
written by new.
memory free function ´
MF
´:
0
3
0
.
5
MF
You can, by activating the memory time function ´
MT
´, display the time that the
memory capacity (see above) has available for recording measured data. This
time depends not only on the memory capacity but also on the number of mea-
suring channels, on the cycle, and even, in certain operating modes, on the
conversion rate.
memory time function ´
MT
´:
0
3
0
.
5
MT
6.5.1 Memory connector
With the ALMEMO 2390-5 it is also possible to connect external ALMEMO
®
EE-
PROM memory connectors ZA 1904-SS with capacities of 128 KB or 256 KB
(25000 or 50000 measured values). These memories do not need a battery to
ensure that data is retained; they can be unplugged, even dispatched elsewhe-
ALMEMO
2390-5
37
Measuring operations