To save the calibration (beam position), press the lock
button on the remote (or the ENT button on the laser
keypad).
If you don’t want to save the calibration, turn the laser off.
To leave the CALIBRATION mode, move the cursor to
EXIT and press ENT.
(If you are using the remote, move up the menu using the
Center Line Button and move down the menu using
the Standby (ZZZ) Button. Press the lock button instead
of ENT. )
To leave the SET UP mode, move the cursor to EXIT and
press ENT.
Final Calibration Check
Re-check your new calibration using the steps shown in the previous section with Stakes A
and B. If the marks are no more than 1/8” apart (at 30m, no more than 3 mm apart), the laser is
within the accuracy spec and calibrated correctly.