9300025700: Rev. A
9
of 14
Pre-installation Instructions
LC/MS
The LC/MS requires the following.
Table 9 LC/MS Circuits, Power Plug and Outlet Types
Component
Circuit
Power Plug and Outlet Type
320-MS
1 each 220V
US: 1 each NEMA 6-15P
EU: 1 each CEE 7/7 plugs
MS40+ Foreline
Pump
1 each 220V
US: 1 each NEMA 6-15P
EU: 1 each CEE 7/7 plugs
Computer
110V or 220V, see appropriate user manual
LC modules
110V or 220V see appropriate user manual
Table 10 LC/MS Power Requirements
Module
Max Current Draw (Amps)
320-MS
10A
MS40+ Foreline Pump
10A
Use a separate dedicated power source for HPLC modules and additional instruments and
equipment. Never plug the mass spectrometer and the chromatograph into the same power
source or the power source may overload. Never use the free outlet on any of the power
sources for equipment that draws more than 2A.
Table 11 MS and Foreline Pump Power Cords
Power Cord
320-MS
MS40+ Foreline Pump
Length
1 each 2.5 m (8 ft)
1 each 2.5 m (8 ft)
Plug
US: 1 each NEMA 6-15P
EU: 1 each CEE 7/7 plugs
US: 1 each NEMA 6-15P
EU: 1 each CEE 7/7 plugs
The power cables for the CPU, monitor, and printer are approximately 2.13 m (7 ft) long.
Figure 3 shows the power plugs and outlet types.
CAUTION
Replacing or substituting power cords or plugs must be done
with strict compliance with all regulations, including electrical
codes, power cord color coding, and appropriate regulatory
agency certification marks.