Troubleshooting the 82357A
Software Configuration Checks
Assigning
usbscan
Driver to the 82357A
If, when you initially plugged an 82357A into a USB port, a Windows Plug
and Play Manager dialog box appeared that requested the location of
usbscan.sys
and you were forced to click
Cancel
from this dialog box,
see the following information in this section. Otherwise, skip this section.
After an initial attachment of an 82357A to a system without
usbscan.sys
installed, future unplugs/replugs of an 82357A will result in the system
(silently) assigning the wrong driver and not querying the user for the
usbscan.sys
driver location. This action leaves the 82357A inoperative,
with all 3 LEDs ON (assuming the Agilent IO Libraries Suite is installed).
To properly associate the
usbscan.sys
driver with an 82357A, you can either
update the driver using the Windows Device Manager utility or uninstall/
reinstall the Agilent IO Libraries Suite (with the 82357A unattached) and
then attach the 82357A and provide the Windows 98 SE media location to
the Plug and Play Manager.
Uninstall/Reinstall the Agilent IO Libraries
With this method, the 82357A install software (found in the Agilent IO
Libraries Suite installer) cleans the Windows Registry of any memory of an
82357A being on the system, forcing the Plug and Play Manager to query
again for the location of
usbscan.sys
. This method has the side effect of
removing any Agilent IO Libraries Suite IO Configurations established
previously.
Disconnect all 82357As from the USB ports. Then, uninstall and reinstall
the Agilent IO Libraries Suite (version L.01.00 or greater, or 14.0 or greater).
Attach the 82357A and provide the Window 98 SE media location to the
Windows Plug and Play Manager.
Update the Driver using Windows Device Manager
With this method, you tell Windows that a new driver is to be assigned to the
82357A. This method requires you to locate where the Device Manager has
mistakenly located the 82357A device.
1
Start Windows Device Manager and select the 82357A device,
usually presented in Device Manager as an “Other” type device.
Then, select Properties for that device and then select the Driver tab
in the resulting dialog.
2
Click the
Update driver...
button and provide the Windows 98 SE CD
when prompted. This forces the Windows Plug and Play Manager to
reassess the drivers needed for that device, allowing you to provide
the
usbscan.sys
source media location.
Содержание USB/GPIB Interface Converter 82357A
Страница 1: ...Agilent 82357A USB GPIB Interface Converter User s Guide ...
Страница 2: ......
Страница 5: ...1 Installing the 82357A ...
Страница 33: ...2 Using the 82357A ...
Страница 42: ...42 Chapter 2 Using the 82357A Setting Configuration Parameters Notes ...
Страница 43: ...3 Troubleshooting the 82357A ...
Страница 57: ...4 82357A User s Guide Information ...
Страница 63: ...A 82357A Specifications ...
Страница 69: ......
Страница 70: ... 82357 90002 User s Guide 82357 90002 Agilent Technologies Inc Printed in Malaysia November 2004 E1104 ...