U2000 Series Operating and Service Guide
73
Table 4-6
Power Sensor SWR and Reflection Coefficient for the U2001H
Table 4-7
Power Sensor SWR and Reflection Coefficient for the U2002H
Table 4-8
Power Sensor SWR and Reflection Coefficient for the U2000B
Table 4-9
Power Sensor SWR and Reflection Coefficient for the U2001B
Frequency
Actual
Measurement
Maximum
Rho
SWR
10 MHz to 6 GHz
0.070
1.15
Frequency
Actual
Measurement
Maximum
Rho
SWR
50 MHz to 8 GHz
0.070
1.15
8 GHz to 12.4 GHz
0.111
1.25
12.4 GHz to 18 GHz
0.123
1.28
18 GHz to 24 GHz
0.130
1.30
Frequency
Actual
Measurement
Maximum
Rho
SWR
10 MHz to 2 GHz
0.057
1.12
2 GHz to 12.4 GHz
0.078
1.17
12.4 GHz to 18 GHz
0.107
1.24
Frequency
Actual
Measurement
Maximum
Rho
SWR
10 MHz to 2 GHz
0.057
1.12
2 GHz to 6 GHz
0.078
1.17