![Agilent Technologies U2000 Series Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/agilent-technologies/u2000-series/u2000-series_operating-and-service-manual_2867112066.webp)
3
Specifications and Characteristics
48
U2000 Series Operating and Service Guide
Power Accuracy
This specification is valid only after zeroing and calibration of the power
sensor at ambient environmental conditions.
Table 3-1
Average only Mode Power Accuracy
1
(with exclusions
)
Table 3-2
Normal Mode Power Accuracy
2
(with exclusions
)
With conditions as follows:
•
After zeroing
3
and 30 minutes of power- on warm- up
•
Number of averages = 1024
Model
Power Level
Accuracy
25 °C
±
10 °C
Accuracy
0 °C to 55 °C
U2000/1/2/4A
–60 dBm to +20 dBm
±
3.0%
±
3.5%
U2000/1/2H
–50 dBm to +30 dBm
±
4.0%
±
5.0%
U2000/1B
–30 dBm to +44 dBm
±
3.5%
±
4.0%
1 The accuracy is essentially a combination of linearity, instrumentation accuracy, and traceability to
absolute accuracy at 50 MHz, 0 dBm. Note: mismatch uncertainty, cal-factor uncertainty, and
power level dependent terms (zero set, drift, and noise) are excluded in this specification and
specified elsewhere in this guide.
Model
Power Level
Accuracy
25 °C
±
10 °C
U2000/1/2A
–30 dBm to +20 dBm
±
4.0%
U2000/1/2H
–20 dBm to +30 dBm
±
5.0%
U2000/1B
0 dBm to +44 dBm
±
4.5%
2 The accuracy for –7 dBm to +1 dBm (U2000/1/2A), +3 dBm to +11 dBm (U2000/1/2H), and +23
dBm to +31 dBm (U2000/1B) power level will be dominated by zero set and measurement noise.
For overall accuracy, refer to the measurement uncertainty calculator which is available on the
Agilent Technologies Web site.
3 It is strongly advisable to perform external zeroing on the U2000 Series USB power sensor for
power measurement level below –30 dBm (U2000/1/2A), –20 dBm (U2000/1/2H), and 0 dBm
(U2000/1B) for accurate measurements. During the external zeroing process, the RF input signal
must be switched off or the device-under-test disconnected from the U2000 Series USB
power sensor.
Содержание U2000 Series
Страница 1: ...Agilent Technologies Agilent U2000 Series USB Power Sensors Operating and Service Guide...
Страница 14: ...xiv U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 16: ...xvi U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 18: ...xviii U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...