Specifications and Characteristics
3
U2000 Series Operating and Service Guide
59
Settling time
In
FAST
mode (using Free Run trigger), for a 10 dB decreasing power
step, the settling time is:
For
Normal
and
x2
speed, manual filter, and a 10 dB decreasing power
step (not across the switching point), refer to
Table 3-5
Settling Time for
Normal
and
x2
Speed
Time
U2000 Series USB Power
Sensors
25 ms
1
1 When a power step crosses the auto-range switch point of
the sensor, add 25 ms.
Number of Averages
1
2
4
8
16
32
64
128
256
512
1,024
Settling Time (s)
(Normal Speed)
0.045
0.09
0.17
0.34
0.66
1.3
2.6
5.2
10.4
20.9
41.9
Settling Time (s)
(x2 Speed)
0.042
0.05
0.09
0.17
0.34
0.66
1.3
2.6
5.2
10.4
20.9