![Agilent Technologies U2000 Series Скачать руководство пользователя страница 89](http://html1.mh-extra.com/html/agilent-technologies/u2000-series/u2000-series_operating-and-service-manual_2867112089.webp)
Service
4
U2000 Series Operating and Service Guide
71
Performance Test
Standing Wave Ratio (SWR) and Reflection Coefficient (Rho)
Performance Test
This section does not establish preset SWR test procedures since there are
several test methods and different equipment available for testing the SWR
or reflection coefficient. Therefore, the actual accuracy of the test
equipment must be accounted for when measuring against instrument
specifications to determine a pass or fail condition. The test system used
must not exceed the system Rho uncertainties shown in the following
tables when testing the U2000 Series USB power sensors.
Table 4-1
Power Sensor SWR and Reflection Coefficient for the U2000A
Table 4-2
Power Sensor SWR and Reflection Coefficient for the U2001A
Frequency
Actual
Measurement
Maximum
Rho
SWR
10 MHz to 30 MHz
0.070
1.15
30 MHz to 2 GHz
0.061
1.13
2 GHz to 14 GHz
0.087
1.19
14 GHz to 16 GHz
0.099
1.22
16 GHz to 18 GHz
0.115
1.26
Frequency
Actual
Measurement
Maximum
Rho
SWR
10 MHz to 30 MHz
0.070
1.15
30 MHz to 2 GHz
0.061
1.13
2 GHz to 6 GHz
0.087
1.19
Содержание U2000 Series
Страница 1: ...Agilent Technologies Agilent U2000 Series USB Power Sensors Operating and Service Guide...
Страница 14: ...xiv U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 16: ...xvi U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 18: ...xviii U2000 Series Operating and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...