62
Setting the Interfaces - SONET Operation
Setting Transmit Interface
patterns are not used for test purposes and are either the same as the test channel
or set to UNEQUIPPED.
Mapping may be selected from a pictorial display by moving the cursor to
MAPPING
and
pressing .
Use
and
to move between STS Layer choice, VT Layer choice and
Payload Layer choice. Use
and
to choose the mapping.
Use
to confirm your choice and return to the
display.
5
If VT-6 mapping is chosen, VT CONCATENATION selection is enabled, choose
or the tributary at which the concatenation begins, VT6-2C through VT6-
6C.
The BACKGROUND, PATTERN IN OTHER VT-6s is fixed at NUMBERED,
that is, each VT-6 contains a unique number to allow identification in case of
routing problems.
6
If FULL SPE, VT-6, VT-2 or VT-1.5 mapping is chosen, choose the test tributary
CHANNEL, including the STS-3 for an OC-12/OC-48 signal.
7
Choose the PATTERN type and PRBS polarity.
SET
SET
SONET MAIN SETTINGS
OFF