Selecting Test Features - SDH Operation
Adding Pointer Adjustments
G.783(g) 87-3 Pattern
An Example of a Pointer Sequence
Pointer Adjustment
87
No Pointer
Adjustment
Start of Next
87-3 Pattern
3
Pointer Sequence
Description
G.783(a) PERIODIC
SINGLE
Periodic Single adjustments, each with opposite polarity to the preceding
adjustment. The interval between pointer adjustments is user selectable (see
Note 1 page 50).
G.783 (b) PERIODIC
ADD
Periodic Single adjustments, with selectable polarity and added adjustment (1
extra). The spacing between the added adjustment and the previous adjustment
is set to the minimum, (see Note 2 page 50). The interval between pointer
adjustments is user selectable (see Note 1). Added adjustments occur every 30
seconds.
G.783 (c) PERIODIC
CANCEL
Periodic Single adjustments, with selectable polarity and cancelled adjustment
(1 less). The interval between pointer adjustments is user selectable (see Note
1 page 50). Cancelled adjustments occur every 30 seconds.
G.783(d) PERIODIC
DOUBLE
Periodic Double adjustments (pair of adjustments). The pair alternate in
polarity. The spacing between pairs of adjustments, of like polarity is set to the
minimum (see Note 2). The interval between pointer adjustments is user
selectable (see Note 1).
G.783 (e) SINGLE
Periodic Single adjustments, all of the same polarity which is selectable.
Separation between pointer adjustments is fixed at approximately 30 seconds.