110
Making Measurements
Performing a Tributary Scan
TIP:
The SDH/SONET Tributary Scan display can be logged to the chosen logging
device. See "Logging on Demand " page 146.
HOW TO:
1
Set up the transmit and receive SDH/SONET interfaces and payload as required.
See “Setting Transmit Interface” on page 20 (for SDH) and page 60 (for SONET)
and also “Setting Receive Interface” on page 30 (for SDH) and page 70 (for
SONET).
2
Choose the required BIT ERROR THRESHOLD.
This determines the error rate above which a failure is declared.
3
Choose the required TEST TIMING.
The value you choose is the test time for each individual tributary and not the total
test time.
For example, 28 VT-1.5 tributaries in an STS-1 SPE - the time taken to complete
the Tributary Scan will be 28 X TEST TIMING choice.
4
The Tributary Scan results can be viewed on the
,
display.
The Scan can be started on the
,
,
display or the
display by choosing START.
If the Scan is started on the
,
,
display, the OmniBER 720 changes to the
display.
RESULTS
SONET TRIBSCAN
TRANSMIT
TEST FUNCTION
SONET TRIB SCAN
RESULTS
TRANSMIT
TEST FUNCTION
SONET TRIB SCAN
RESULTS