Calibrate Commands
CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept any calibration
commands. The first parameter specifies the state. The second optional parameter is the password.
Parameter
Typical Return
0|OFF|1|ON , default OFF
0 or 1
<
password
> a numeric value up to 15 digits
(none)
Disable calibration mode and unsecure calibration: CAL:STAT OFF,0
Enable calibration: CAL:STAT ON
<
password
> is optional to secure the instrument, but must be correct if provided.
CALibrate:VOLTage[:LOCal][:LEVel] <
value
>
Calibrates the local voltage programming and measurement. The value selects the range to calibrate.
Parameter
Typical Return
The maximum voltage of the range
being calibrated.
(none)
Calibrates the voltage of the 80 V range: CAL:VOLT 80
CALibrate:VOLTage:REMote[:LEVel] <
value
>
Calibrates the remote voltage programming and measurement. The value selects the range to calibrate.
Parameter
Typical Return
The maximum voltage of the range
being calibrated.
(none)
Calibrates the voltage of the 80 V range: CAL:VOLT:REM 80
Agilent N8900 Series Operating and Service Guide
79
Содержание N8900 series
Страница 1: ...Agilent N8900 Series Autoranging System DC Power Supply Operating and Service Guide...
Страница 36: ...Multiple Unit Connections Agilent N8900 Series Operating and Service Guide 35...
Страница 104: ...Status Tutorial Status Diagram Agilent N8900 Series Operating and Service Guide 103...
Страница 117: ...Service and Maintenance Service and Maintenance Pending 116 Agilent N8900 Series Operating and Service Guide...
Страница 118: ...Verification and Calibration Verification and Calibration Pending Agilent N8900 Series Operating and Service Guide 117...
Страница 119: ......