Parallel Operation
Front Panel Menu
SCPI Command
Select
System\Group\Master
The Configured field displays the addresses of
all connected and configured Slave units.
If the Configured field is blank, press Dis-
cover Slaves to discover and display the
addresses of the connected units.
Not available
Group Protection
The following protection features are implemented in a grouped configuration:
Power On
At power-on, all grouped units will have their outputs off (unless they have
been configured by the
*RCL0
command to be programmed on).
Configuration
check
At power on, the master unit will search for the slave units and compare the list
of slaves found with the previously stored configuration. If the lists match, nor-
mal operation continues. If the lists do not match, a Master/Slave Protection
(MSP) fault occurs and the outputs of all units remain off (see
Programming
Output Protection
). A message is placed in the Error Queue that lists the slave
units found and the slave units expected.
Com-
munications
check
If a master unit loses communication with any slave at any time, it will gen-
erate an MSP fault, which turns its own output off as well as the outputs of the
slaves that it can still communicate with. If a slave loses communication with
the master, the slave will also turn its output off.
Agilent N8900 Series Operating and Service Guide
61
Содержание N8900 series
Страница 1: ...Agilent N8900 Series Autoranging System DC Power Supply Operating and Service Guide...
Страница 36: ...Multiple Unit Connections Agilent N8900 Series Operating and Service Guide 35...
Страница 104: ...Status Tutorial Status Diagram Agilent N8900 Series Operating and Service Guide 103...
Страница 117: ...Service and Maintenance Service and Maintenance Pending 116 Agilent N8900 Series Operating and Service Guide...
Страница 118: ...Verification and Calibration Verification and Calibration Pending Agilent N8900 Series Operating and Service Guide 117...
Страница 119: ......