12
Chapter 1
Installation
Operation Verification
4.
On the signal generator, select the device interface test:
Press
Aux Fctn
>
N5102A Interface
>
Diagnostics
>
Loop Back Test Type
>
Device Intfc
.
As shown in
Figure 1-6
, the currently selected test is displayed in parenthesis below the
Loop Back Test Type
softkey. Note also that the graphic provided displays the current test setup.
Figure 1-6
ESG/PSG Diagnostic Test Display
5.
Run the selected test: press
Run Loop Back Test
.
When the test completes, the results of the test (pass or fail) replaces
Unknown
in both the
parenthesis with the softkey and in the
Status Summary
display as shown in
Figure 1-7
. If
this test fails, refer to
“Running Diagnostic Tests” on page 108
.
Figure 1-7
ESG/PSG Loop Back Test Result
Selected loop back test
Loop back test setup
Loop back test results