Chapter 6
107
Troubleshooting
If You Encounter a Problem
Checking Power Problems
1.
Check the power cord; ensure that it is:
•
in good condition
•
properly plugged in to a live outlet (line power connection is described on
page 6
)
•
properly connected to the power supply (power supply connection is described on
page 10
) and the
DC power supply plug is fully inserted into the N5102A module DC power receptacle
If this does not solve the problem, go to step 2 to check the power supply.
2.
Using a DVM, check the power supply output.
When you connect the power supply to the module, the green Power LED should light.
If the Power LED does not light:
If the output voltage is not correct, replace the power supply
(refer to
“Replaceable Parts” on page 112
).
If the output voltage is correct, the problem is the N5102A
module. Return the module to Agilent (see
page 113
).
+4.7 to 5.3V