Getting Started
14
Clock Synthesizer 13.5 GHz User Guide
•
Connect the
Trig-O
to the trigger input of your sampling instrument (the
default trigger output frequency is clock frequency/8).
•
If using an external reference, connect the reference to the
10 MHZ IN
BNC connector.
1.4
Setting Frequency and Output Levels
•
Turn on the power switch from the back panel.
•
The N4963A synthesizer starts with the
FREQ
option selected by default
and "005.000G" showing in the display.
•
Set the desired frequency using the
Unit, Digit
and
Value
buttons.
•
Press the
Display-
>
Scroll
button and select
AMP.
Select the output
channel bank with the
Chan
button. Set the desired single-ended peak-to-
peak voltage using the
Unit, Digit
and
Value
buttons.
•
Press the
Display-
>
Scroll
button and select
DC OFFSET
. Select the
output channel bank with the
Chan
button. Set the desired DC offset
voltage using the
Digit
and
Value
buttons.
•
If using an external reference, press the
Ext 10MHz
button, and the
indicator will light. The internal reference is being used if the indicator is
off.
•
To turn on the clock output and the trigger signal, press the
Clock
->
On
button. The green LED will light indicating the synthesizer is transmitting.
1.5
Aligning Clock and Data
•
The front panel
Reference Output
(channel 1) signal can be phase shifted
relative to all other output channels.
•
Press the
Display-
>
Scroll
button and select
DELAY
. Set the phase
offset between channel 1 and all other channels using the
Digit
and
Value
buttons.
•
Center the reference clock for minimum BER or other desired state.