Check the Output Signal (Clock Multiplier Function of N4916B)

The general approach shown here is using an oscilloscope.
 If you have connected the N4916B to a High-Performance 
Serial BERT N4903B, you can also use the error detector. 
For details see the Help and the User Guide.

Choose a suitable pattern 
(e.g. PRBS) and start the 
pattern generator.

8

After some adjustments of an

, you will see a 

display like the following:

oscilloscope

9

In case of problems, please refer to the section 

"Problems with the N4916B" of the Help and the 

User Guide of your N4903B.

Preset the instrument state.

1

Preset the instrument state.

1

Enable Clock Multiplier 
from 

Config

 window.

4

In the 

Clock Multiplier DATAOUT

 window,

set the values as following:
- Input frequency to 3 Ghz
- Multiplier to 2
- Termination ON at 0 V

5

In the 

PG-Bit Rate Setup

 window,

set Datarate to 3 Gb/s.

2

In the 

PG-Clock/Trigger Output

 window,

set the Clock Offset to 0 V,  
Clock Out Amplitude to 400 mV 
and Trigger Out Offset to 0 V.

3

Connect the Clock Out of
PG to the Clock In of 
Clock Multiplier. 

6

Connect the Clock Out of 
Clock Multiplier to an

oscilloscope

7

Содержание N4916B

Страница 1: ...nline Help of your High Performance Serial BERT N4903B For more information visit The Help also offers printable versions of the User Guide and the Programming Guide www agilent com find jbert You onl...

Страница 2: ...I pollution degree 2 environment It is designed to operate within a temperature range of 5 40 C 40 105 F at a maximum relative humidity of 95 and at altitudes of up to 2000 meters Refer to the specifi...

Страница 3: ...the installation media All data and clock input and output connectors of the PG that are not used in your test setup must be terminated with 50 Ohm The instrument can operate from any single phase AC...

Страница 4: ...ct the Agilent IO Control icon in the Windows task bar 5 If you see this icon the revision of the Agilent IO Libraries Suite is 15 5 or later You may wish to connect a keyboard and mouse to your Seria...

Страница 5: ...he rear USB ports of your N4903B 3 When you connect the N4916B for the first time the Windows Found New Hardware Wizard appears and asks whether it should connect to Windows Update Answer No 4 Click N...

Страница 6: ...e alias name to DATA_OUT or AUXDATA_OUT depending on the channel Data Out and Aux Data Out used This is the alias used by the software 3 After clicking OK the Agilent Connection Expert shows a window...

Страница 7: ...these functions 5 Enabling Clock Multiplier Select the Clock Multiplier function to open the Clock Multiplier Connection dialog 7 If you select the Clock Multiplier function a new menu entry is added...

Страница 8: ...For details see the Help and the User Guide Choose a suitable pattern e g PRBS and start the pattern generator 9 After some adjustments of the scope you will see a display like the following 10 In ca...

Страница 9: ...of problems please refer to the section Problems with the N4916B of the Help and the User Guide of your N4903B Preset the instrument state 1 Preset the instrument state 1 Enable Clock Multiplier from...

Отзывы: